WHITESTAR SIGNATURE PRO System 3 • Surgery Start Up Review Case Figure 3.20 – Review Case Screen Review Case – shows the graphical history of the surgery. (Refer to Figure 3.20 – Review Case Screen.) 1. Select a shorter time interval to adjust the time line duration and to see greater case details. For example, 10 Sec shows more detail than 15 Min. 2. Select a longer time interval for an overview of the surgery or sessions, such as 1 hour or 2 hours.
4 SURGEON SETTINGS Select Surgeon Add a New Surgeon Delete Surgeon Edit a Surgeon Foot Pedal Setup Set Sound Levels Surgeon – Assign Order RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings When you use the system for the first time, Standard Surgeon is the only surgeon on the system. Select Surgeon RX Only- NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System Add a New Surgeon 4 • Surgeon Settings To create a new surgeon: 1. 2. 3. 4. Press on Select Surgeon. The keyboard screen opens. Enter the name of the surgeon. Press Enter to save the new surgeon and then exit the keyboard screen. The new surgeon is now on the list of available surgeons. Figure 4.2 – Keyboard Screen Delete Surgeon To remove a surgeon name from the list: 1. Select the name of the surgeon. 2. Press on Select Surgeon. 3.
WHITESTAR SIGNATURE PRO System Edit a Surgeon 4 • Surgeon Settings 1. Select a surgeon from the list. Figure 4.3 – Select Surgeon 2. Press on Select Surgeon. 3. The keyboard screen appears. RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Figure 4.4 – Edit Surgeon Name Screen 4. Edit the name. 5. Press Enter. RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System Foot Pedal Setup 4 • Surgeon Settings This section presents the setup, calibration, and testing process for your foot pedal, divided into three sections: • Single Linear Foot Pedal • Advanced Control Pedal • Advanced Linear Pedal Single Linear Foot Pedal Setup This section presents the following topics for setting up a standard foot pedal.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Table 4.1 – Single Linear Foot Pedal Default Settings Parameter Pitch regions Available Settings 1–100% Default Settings P1=5% P2=30% P3=60% Switch position Switch1= * Switch 2= * Switch1=Off Switch 2=Off Right= * Left= * Right=Reflux Left=Off On, Off Off Feedback *Table 4.3 – Single Linear Foot Pedal Switch Assignments Vitrectomy Foot Pedal Modes Table 4.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Single Linear Foot Pedal Connections Attach the foot pedal cable to the connector on the left side of the foot pedal and attach the other end of the cable to the foot pedal connector on the rear panel of the system. Figure 4.6 – Rear Panel Connections 1. USB port 2. Communication port 3. Foot pedal connector 4. Compressed air 5. Potential equalizer 6. Power switch RX Only – NGP Z370500 Rev. B 1116 7. Fuse holder 8. Power cord connector 9.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Single Linear Foot Pedal Switch Assignment The standard foot pedal has four switches that you can assign specific functions for each surgeon. For Phaco and Vitrectomy foot pedal settings the switches are set per submode. Follow these steps to assign functions to the foot pedal switches. Note: In addition to the procedure below, you can access the Single Linear Foot Pedal Summary screen by pressing the foot pedal symbol near the top of the screen. 1.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Figure 4.8 – Switches Screen - Single Linear Foot Pedal 5. Press the Switches button to change the settings for the switches. The system displays the standard foot pedal switch assignment screen. The Single Linear foot pedal Switches screen allows you to assign or change the action for each of the four switches. 6. Select the major mode and then the applicable submode. 7.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Figure 4.9 – Single Linear Foot Pedal Switch Assignment Dialog Box 8. Press the button that represents your choice of switch setting. Table 4.3 – Single Linear Foot Pedal Switch Assignments Switch Setting Off Description This switch has no switch assignment. Surgeon Program Up Selects the surgeon program above the current program in the list. Surgeon Program Down Selects the surgeon program below the current program in the list.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Switch Setting Previous Sub Mode Description Changes to the previous submode. Next Sub Mode Changes to the next submode. Toggle SMC Record Activates or deactivates the surgical media center (SMC) recording function. Bottle Up Raises the IV pole while the switch is held down. Bottle Down Lowers the IV pole while the switch is held down. Reflux Activates or deactivates reflux.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Figure 4.10 – Single Linear Foot Pedal Thresholds 4. Select the Feedback box to enable foot pedal feedback. Feedback is a physical and audible click as you move the foot pedal to the next or previous position. 5. Press Save to retain the settings. RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Single Linear Foot Pedal Calibration and Testing Follow these steps to calibrate and test your foot pedal. 1. On the Surgeons and Programs screen, press 2. Select the Calibration button. Note: . The Calibration feature is not available during surgery. Figure 4.11 – Single Linear Foot Pedal Calibration Screen 3. With your foot off the foot pedal, press the Phase 1 button. 4.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Figure 4.12 – Single Linear Foot Pedal Testing Screen 9. To test the foot pedal before calibration, press the Test button. The Test button is not available during surgery. RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Advanced Control Pedal Setup This section presents the following topics for setting up an Advanced Control Pedal (ACP): • Advanced Control Pedal Connections • Advanced Control Pedal Switch Assignment • Advanced Control Pedal Threshold Configuration • Advanced Control Pedal Calibration and Testing Figure 4.13 – Advanced Control Pedal 1. Left yaw 2. Wake-up button 3. Left heel switch Note: 4. Heel rest 5. Right heel switch 6. Lights 7. Right yaw 8.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Vitrectomy Foot Pedal Modes Table 4.5 – Vitrectomy Advanced Control Pedal Modes Foot Pedal Positions Vitrectomy Foot Pedal Modes ICA IAC Side VIT 1 Irrigation Irrigation N/A 2 Irrigation/Cut Irrigation/ Aspiration N/A 3 Aspiration Cut N/A Advanced Control Pedal Connections You can attach the Advanced Control Pedal to the system either with a cable or with a wireless BLUETOOTH connection.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Figure 4.14 – Advanced Control Pedal Phaco Summary Screen 3. Press the Switches button to change the settings for the switches. The system displays the Advanced Control Pedal edit screen. The Advanced Control Pedal edit screen allows you to assign or change the action for each of the four switches. RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Figure 4.15 – Advanced Control Pedal Phaco Switches Screen 4. Select the major mode and then the applicable submode. 5. Press the button under each switch name to assign or change the action for that switch. The system displays a switch assignment dialog box with the text of the current switch assignment shown in green.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Figure 4.16 – Advanced Control Pedal Switch Assignment Dialog Box 6. Press the button that represents your choice of switch setting. Table 4.6 – Advanced Control Pedal Switch Assignments Switch Setting Off Description This switch has no switch assignment. Surgeon Program Up Selects the surgeon program above the current program in the list. Surgeon Program Down Selects the surgeon program below the current program in the list.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Switch Setting Bottle Up Description Raises the IV pole while the switch is held down. Bottle Down Lowers the IV pole while the switch is held down. Reflux Activates or deactivates reflux. Continuous Irrigation Activates or deactivates continuous irrigation. 1-Touch Up Increments CASE mode 1-Touch. (Phaco only) 1-Touch Down Decrements CASE mode 1-Touch.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Advanced Control Pedal Threshold Configuration Follow these steps to configure the thresholds for the foot pedal. Note: In addition to the procedure below, you can access the Advanced Control Pedal Summary screen by pressing the foot pedal symbol near the top of the screen. 1. On the Surgeons and Programs screen, press . 2. Press the Thresholds button under Foot Pedal. The system displays the foot pedal thresholds screen. Figure 4.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Advanced Control Pedal Calibration and Testing Follow these steps to calibrate and test your foot pedal. Note: In addition to the procedure below, you can access the Advanced Control Pedal Summary screen by pressing the foot pedal symbol near the top of the screen. 1. On the Surgeons and Programs screen, press 2. Press the Calibration button. . Figure 4.18 – Advanced Control Pedal Calibration 3.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Figure 4.19 – Advanced Control Pedal Testing 7. Press the foot pedal down fully to position 3. As the foot pedal travels through positions 1 and 2 to position 3, the progress bar shows its current state. When the pedal reaches the bottom of position 3, the progress bar should be solid green. 8. Press each of the four switches. When a switch is pressed, the On button for that switch turns green and its Off button turns white.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Advanced Linear Pedal Setup Note: Only use the Abbott Medical Optics specified replacement battery in the Advanced Linear Pedal to ensure correct system performance. See “Charging Options for Wireless Devices” on page 8-5. The Advanced Linear Pedal does not need to be calibrated. Figure 4.20 – Advanced Linear Pedal 1. Handle 2. Left and Right Top Switches 3. Left and Right Side Switches 4. Heel rest 5.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Advanced Linear Pedal Switch Assignment The Advanced Linear Pedal has four switches that you can assign specific functions to. Follow these steps to assign functions to the foot pedal switches. Note: In addition to the procedure below, you can access the Advanced Linear Pedal Summary screen by pressing the foot pedal symbol near the top of the screen. 1. On the Surgeons and Programs screen, press Note: .
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Figure 4.22 – Advanced Linear Pedal Switches Screen 3. Select the major mode and then the applicable submode. 4. Press the button under each switch to assign or change the action for that switch. The system displays a switch assignment dialog box with the text of the current switch assignment shown in green.