WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Figure 4.23 – Advanced Linear Pedal Switch Assignment Dialog Box 5. Press the button that represents your choice of switch setting. Table 4.7 – Advanced Linear Pedal Switch Assignments Switch Setting Off Description This switch has no switch assignment. Surgeon Program Up Selects the surgeon program above the current program in the list. Surgeon Program Down Selects the surgeon program below the current program in the list.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Switch Setting Bottle Up Description Raises the IV pole while the switch is held down. Bottle Down Lowers the IV pole while the switch is held down. Reflux Activates or deactivates reflux. Continuous Irrigation Activates or deactivates continuous irrigation. 1-Touch Up Increments CASE mode 1-Touch (phaco only). 1-Touch Down Decrements CASE mode 1-Touch (phaco only). Toggle CASE Activates or deactivates CASE mode, (phaco only) 6.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Advanced Linear Threshold Configuration RX Only- NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Follow these steps to set the foot pedal thresholds. Note: In addition to the procedure below, you can access the Advanced Linear Pedal Summary screen by pressing the foot pedal symbol near the top of the screen. 1. On the Surgeons and Programs screen, press . 2. Press the Thresholds button under Foot Pedal. The system displays the foot pedal thresholds screen. 3.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Advanced Linear Pedal Test Figure 4.25 – Advanced Linear Pedal Test Screen Note: In addition to the procedure below, you can access the Advanced Linear Pedal Summary screen by pressing the foot pedal symbol near the top of the screen. 1. On the Surgeons and Programs screen, press . 2. Press the Test button. 3. Press the foot pedal down fully to position 3.
WHITESTAR SIGNATURE PRO System Set Sound Levels 4 • Surgeon Settings Default Settings Table 4.
WHITESTAR SIGNATURE PRO System 4 • Surgeon Settings Follow these steps to set the system sound levels. 1. On the Surgeons and Programs screen, select the surgeon. 2. Press . Figure 4.26 – Sound Settings Screen 3. Use the sliders to adjust the volume for: • Instrument • Confirmation 4. Use the buttons on the right side to turn sounds on or off for the different system functions. 5.
WHITESTAR SIGNATURE PRO System Surgeon – Assign Order 4 • Surgeon Settings 1. Select a surgeon. 2. Use or to move the surgeon names up or down on the Surgeons and Programs screen. Figure 4.27 – Surgeons and Programs Screen 3. Repeat steps until the order is to your satisfaction. RX Only – NGP Z370500 Rev.
5 PROGRAM SETTINGS Create a New Program Delete a Program Edit a Program Name Program Settings Phaco Irrigation and Aspiration Vitrectomy Diathermy Default Settings Phaco Power Settings WHITESTAR Technology FUSION Fluidics Phaco Occlusion Mode Settings CASE Mode Passive Reflux Lock a Program Copy a Surgeon Program Program – Assign Order RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System Create a New Program 5 • Program Settings When you create a new surgeon, the system automatically assigns the Default Anterior Program and the Post Laser Cataract to that surgeon. You can customize a program for that surgeon name with your preferred settings. You can also create new programs for existing surgeons. Figure 5.1 – Surgeons and Programs Screen 1. 2. 3. 4. 5. Delete a Program On Select Program, press .
WHITESTAR SIGNATURE PRO System 5 • Program Settings Figure 5.2 – Delete Program Confirmation Dialog Box 3. Press Yes. Edit a Program Name 1. Select a program name on the list. 2. Press . 3. Enter a different program name or select a new name from the Standard Names list. 4. Press Standard Names. 5. Select a program from the list. 6. Press Enter. Program Settings Program settings are used to define your specific settings on the WHITESTAR SIGNATURE PRO system.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Phaco 1. Press to open the Settings screen. Figure 5.3 – Phaco Settings Screen 2. 3. 4. 5. Select a submode or customize the submode name. Select the pump type (peristaltic or Venturi). Use the up and down arrows to set the Aspiration Rate (10 to 60). Use the up and down arrows to set the Vacuum. (0 to 650) If you are using Venturi, the range is 0 to 600. 6. Use the up and down arrows to set the Power (0% to 100%). 7.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Irrigation and Aspiration Figure 5.4 – Irrigation/Aspiration Settings 1. 2. 3. 4. Select a submode or customize the submode name. Select the pump type (peristaltic or Venturi). Use the up and down arrows to set the Aspiration Rate (0 to 60). Use the up and down arrows to set the Vacuum. (0 to 650) If you are using Venturi, the range is 0 to 600. 5. Select the delivery mode (Linear or Panel). 6. Set the bottle height 7. Press Save .
WHITESTAR SIGNATURE PRO System 5 • Program Settings Vitrectomy Figure 5.5 – Vitrectomy Settings Screen 1. 2. 3. 4. Select a submode or customize the submode name. Select the pump type (peristaltic or Venturi). Use the up and down arrows to set the Aspiration Rate (0 to 60). Use the up and down arrows to set the Vacuum. (0 to 650) If you are using Venturi, the range is 0 to 600. 5. Use the up and down arrows to set the Cut Rate (50 to 2500).
WHITESTAR SIGNATURE PRO System 5 • Program Settings Diathermy Figure 5.6 – Diathermy Power Submode 1. Select a submode or customize the submode name. 2. Use the up and down arrows to set the Power (5% to 100%). 3. Select the delivery mode (Linear, Panel, or Burst). 4. Press Save . RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Default Settings Table 5.
WHITESTAR SIGNATURE PRO System Parameter Available Settings 5 • Program Settings Default Settings Max power Unoccluded (%) 40 20 5 30 Power delivery Continuous Continuous Continuous Continuous WHITESTAR setting (duty cycle) On 6/12 (33%) On 6/12 (33%) On 6/12 (33%) On 6/12 (33%) 40 30 10 40 Power delivery Continuous Continuous Continuous Continuous WHITESTAR setting (duty cycle) On 6/12 (33%) On 6/12 (33%) On 6/12 (33%) On 6/12 (33%) Power type, occluded and unoccluded Linea
WHITESTAR SIGNATURE PRO System 5 • Program Settings Table 5.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Table 5.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Table 5.
WHITESTAR SIGNATURE PRO System Phaco Power Settings 5 • Program Settings • Continuous • Pulse Mode • Short Pulse • Long Pulse • Low Power Pulse • High Power Pulse • Single Burst (panel only) • Multiple Burst (panel only) • Continuous Burst From the Phaco Settings screen press to show the Phaco Power selection. Figure 5.7 – Phaco Power Settings In addition to the nine phaco power modes, you can also select the WHITESTAR mode. Additional WHITESTAR Technology information is later in this section.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Continuous Phaco Power Continuous phaco power delivers continuous, uninterrupted phaco power to the handpiece and requires no pulse rate setting. Pulse Mode Pulse Mode delivers phaco in pulses of 1 to 100 ms when the foot pedal is in position 3. You can set this in a range of 1 to 100 pulses per second (pps). The default setting is 20 pps. To set the Pulse Mode range: 1. 2. 3. 4. Press the Pulse Mode button. Press on the pulse setting number.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Single Burst Phaco Power (Panel Only) Single Burst delivers a single burst of ultrasonic power of 110 ms duration when you press the foot pedal to position 3. You must return to foot pedal position 2, pause for approximately one-half (0.5) second, and then press the foot pedal to position 3 to obtain an additional burst of energy.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Table 5.6 – WHITESTAR Technology Parameter Settings Setting Pulse Off Time (ms) 12 Duty Cycle 6/12 Pulse On Time (ms) 6 33% Pulse Rate (pps) 55 4/8 4 8 33% 83 6/4 6 4 60% 100 6/8 6 8 43% 71 8/4 8 4 67% 83 4/24 4 24 14% 35 6/28 6 28 18% 29 6/24 6 24 20% 33 8/24 8 24 25% 31 6/18 6 18 25% 41 Figure 5.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Figure 5.9 – Variable WHITESTAR Technology Foot Pedal Positions and Duty Cycles 1. Foot pedal position 3 2. WHITESTAR duty cycle Figure 5.10 – Duty Cycles for Variable WS RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 5 • Program Settings WHITESTAR ICE Technology (Pulse Shaping) The WHITESTAR ICE Technology was the next micro-pulse advance in phacoemulsification technology, which combined modulated ultrasonic power (pulse shaping) with vacuum control through the application of the Chamber Stabilization Environment (CASE).
WHITESTAR SIGNATURE PRO System 5 • Program Settings Figure 5.12 – Pulse Shaping Screen There are four settings for WHITESTAR ICE Pulse Shaping: • Low Power Limit • High Power Limit • Percent Kick Low End of Range • Percent Kick High End of Range The Low Power Limit and High Power Limit settings define the range of the applied pulse shaping. When the applied phaco power is outside these limits, there is no pulse shaping.
WHITESTAR SIGNATURE PRO System FUSION Fluidics Phaco 5 • Program Settings FUSION Fluidics Phaco is an intelligent vacuum monitoring system that regulates the maximum allowable vacuum that follows an occlusion of the phaco tip. When the phaco tip becomes occluded, the vacuum rises. Clearing of the occlusion while the vacuum is at a high-level can cause a post-occlusion surge.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Figure 5.14 – Phaco Surgical Screen with Occlusion and CASE On Occlusion Mode Settings With Occlusion Mode, you can set different aspiration rates for occluded aspiration as opposed to unoccluded aspiration. You can set a different vacuum rise time for when the phaco tip occludes without changing the aspiration rate through an unoccluded needle. To set the occluded aspiration rate thresholds: 1. Press Phaco. 2. Press Settings. 3.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Figure 5.15 – Occlusion Mode Settings Screen 4. Use the up and down arrows to adjust the pump ramp ). 5. Press the FUSION button ( 6. On the FUSION screen, Use the up and down arrows to adjust the threshold and the Actual Max Vac. RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Figure 5.16 – FUSION Settings Screen 7. Press Finished to close the screen. 8. Press Save to retain the settings. 9. Press Exit to close the screen. RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 5 • Program Settings Occlusion Vacuum Threshold In Occlusion mode phaco, you can set an occlusion threshold value for vacuum. When in Occlusion mode phaco, there is an additional control panel for vacuum. The vacuum threshold setting lets you choose the vacuum level at which occluded settings take effect. To adjust the occluded vacuum threshold, press the up or down arrows. When you decrease the threshold, the occlusion settings take effect sooner.