WHITESTAR SIGNATURE PRO System 8 • Error Messages Troubleshooting and Diagnostics 513 IV Pole communications error. Probable cause: Communication error. Corrective action: 1. Prime the system. 2. If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service. 514 IV Pole error. Probable cause: Communication error. Corrective action: 1. Prime the system. 2.
WHITESTAR SIGNATURE PRO System 8 • Error Messages Troubleshooting and Diagnostics 517 IV Pole is jammed. Probable cause: • IV Pole motor failure. • There is an obstruction that does not allow the IV pole to move. Corrective action: 1. Check the IV pole for obstructions. 2. Prime the system. 3. If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service. 601 Tune excessive vacuum error. Probable cause: There is too much vacuum.
9 WARRANTY AND MAINTENANCE Warranty Statement RX Only –NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System Warranty Statement 9 • Warranty and Maintenance Abbott Medical Optics Inc. warrants for a period of one year (12 months) from the date of installation of the WHITESTAR SIGNATURE PRO System console, foot pedal, wireless remote control, the programmable power pole, and the phaco handpiece to be free from defects in materials and workmanship when properly installed, maintained, and used for the intended purpose.
WHITESTAR SIGNATURE PRO System 9 • Warranty and Maintenance Maintenance User maintenance of the WHITESTAR SIGNATURE PRO System is only for those adjustments and corrective actions in the Error Messages and Troubleshooting, and Diagnostics sections of this manual. There are no user serviceable components within the console and you must not attempt to access the internal components. Any attempt to do so will void the warranty.
10 SPECIFICATIONS Physical Specifications Environmental Specifications Electrical Specifications Phacoemulsification Specifications Irrigation and Aspiration Specifications Vitrectomy Specifications Diathermy Specifications Diathermy Power Graphs Diathermy Power versus Load Impedance Phaco Power Graphs Phaco Tip Velocity Bottle Height Pressure RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System Physical Specifications 10 • Specifications Table 10.1 – System Physical Specifications English / U.S. Metric: Width 24 inches 61 cm Depth 24 inches 61 cm Height 59 inches 150 cm Power cord length 20 feet 6 meters Electrical enclosure current leakage IEC 60601-1 compliance UL 60601-1 Weight (including IV pole) 185 pounds 84 kg Width 12 inches 31 cm Length 10.5 inches 27 cm Height 5.5 inches 14 cm 10.0 pounds 4.5 kg Width 10.
WHITESTAR SIGNATURE PRO System 10 • Specifications English / U.S. Metric: Width 5 inches 13 cm Length 5 inches 13 cm Height 1.5 inches 4 cm Weight 2.0 pounds .9 kg Wireless Remote Control Environmental Specifications Table 10.
WHITESTAR SIGNATURE PRO System Irrigation and Aspiration Specifications 10 • Specifications Table 10.5 – Irrigation and Aspiration Specifications Peristaltic Pump Venturi Pump Vitrectomy Specifications Vacuum Level 0–650 mmHg in 5 mmHg increments Pump Flow 0-60 cc/minute Vacuum Level 0-600 mmHg in 5 mmHg increments Table 10.
WHITESTAR SIGNATURE PRO System Diathermy Specifications 10 • Specifications Table 10.7 – Diathermy Specifications Power Adjustment 5% to 100%, in 5% increments Diathermy Power (maximum power into rated load) 8.5 Watts into 350 Ohms 386 kHz Diathermy Type Bipolar Rated Peak Voltage 160 Volts Connector of diathermy forcep/ probe Ø.072 RECEPTACLE (2) All measurements are in inches. RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 10 • Specifications Diathermy Power Graphs Figure 10.1 – Diathermy Output Power (Typical) 10 Ohm Y Axis = Watts X Axis = Settings Figure 10.2 – Diathermy Output Power (Typical) 50 Ohm Y Axis = Watts X Axis = Settings RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 10 • Specifications Figure 10.3 – Diathermy Output Power (Typical) 200 Ohm Y Axis = Watts X Axis = Settings Figure 10.4 – Diathermy Output Power (Typical) 350 Ohm Y Axis = Watts X Axis = Settings RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 10 • Specifications Figure 10.5 – Diathermy Output Power (Typical) 500 Ohm Y Axis = Watts X Axis = Settings Figure 10.6 – Diathermy Output Power (Typical) 1000 Ohm Y Axis = Watts X Axis = Settings RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System Diathermy Power versus Load Impedance 10 • Specifications Figure 10.7 – Diathermy Output Power (50% Setting) Y Axis = Watts X Axis = Impedance (Ohms) Figure 10.8 – Diathermy Output Power (100% Setting) Y Axis = Watts X Axis = Impedance (Ohms) RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System Phaco Power Graphs 10 • Specifications Figure 10.9 – Longitudinal Phaco Power Y Axis = Output Power (Watts) X Axis = GUI Settings % Figure 10.10 – ELLIPS FX Phaco Power Linear Mode Y Axis = Power (Watts) X Axis = GUI Settings % RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System Phaco Tip Velocity Bottle Height Pressure 10 • Specifications Nominal Velocity for the ELLIPS FX handpiece and the WHITESTAR handpiece at 100% power, continuous, with the OPOR3020L phaco tip is shown in the following table. Handpiece Nominal Longitudinal Velocity (m/s) ELLIPS FX (690880 (A)) 8.85 WHITESTAR (690697 (A)) 11.
WHITESTAR SIGNATURE PRO System 10 • Specifications An IV bottle height reported as 0 cm includes an offset of 7.9 inches or 20 cm to make sure that a positive pressure is always provided. Table 10.8 – Expected Pressure as a function of Bottle Height Bottle Height cm Theoretical Pressure mmHg Corrected Pressure At Patient Eye Level mmHg 0 0.00 14.80 20 14.80 29.60 40 29.60 44.40 60 44.40 59.20 80 59.20 74.00 100 74.00 88.80 105 77.70 92.50 RX Only – NGP Z370500 Rev.
11 ACCESSORIES AND PARTS REORDERING Accessories and Parts List RX Only – NGP Z370500 Rev.
WHITESTAR SIGNATURE PRO System 11 • Accessories and Parts Reordering Table 11.1 – WHITESTAR SIGNATURE PRO System Part Number Name WHITESTAR SIGNATURE PRO Console List of Accessories with Part Numbers Description An ophthalmic microsurgical system. Part number NGP680301 The following table is a list of accessories that you can use to reorder parts for your WHITESTAR SIGNATURE PRO System.
WHITESTAR SIGNATURE PRO System System Accessories Diathermy Handpiece Vitrectomy Cutter 11 • Accessories and Parts Reordering Part Number Description OM05510112 PHACOFIT O-Ring's I/A Tip (5PK) OM05510113 PHACOFIT Titanium, Multi-Tip Set OM05510114 PHACOFIT Cleaning/Flushing Kit OM05510115 PHACOFIT Tip, Luer Adapter (Aspiration) OM05510116 PHACOFIT Tip, 45 Degree Silicone (No Sleeve) OM0551011 PHACOFIT I/A Tip Straight 0.3 MM Port OM0551012 PHACOFIT I/A Tip Curved 0.
WHITESTAR SIGNATURE PRO System System Accessories FUSION Pack 11 • Accessories and Parts Reordering Part Number Description OPO70 FUSION Pack (disposable, sterile) OPO71 FUSION Dual Pump Pack (disposable, sterile) Wireless Remote Control NGP680137 WHITESTAR SIGNATURE PRO System Remote Control 0100-1834 AC Charger, SIGNATURE Wireless Remote Foot Pedal NGP680701 WHITESTAR SIGNATURE System Single Linear Foot Pedal NGP680702 WHITESTAR SIGNATURE System Advanced Control Pedal NGP680703 WHITESTA
WHITESTAR SIGNATURE PRO System System Accessories 11 • Accessories and Parts Reordering Part Number Description NGP Z370504 WHITESTAR SIGNATURE PRO System Operator’s Manual (German) NGP Z370505 WHITESTAR SIGNATURE PRO System Operator’s Manual (Greek) NGP Z370506 WHITESTAR SIGNATURE PRO System Operator’s Manual (Hungarian) NGP Z370507 WHITESTAR SIGNATURE PRO System Operator’s Manual (Italian) NGP Z370508 WHITESTAR SIGNATURE PRO System Operator’s Manual (Japanese) NGP Z370509 WHITESTAR SIGNATUR
WHITESTAR SIGNATURE PRO System System Accessories 11 • Accessories and Parts Reordering Part Number Description NGP Z370526 WHITESTAR SIGNATURE PRO System Operator’s Manual (Traditional Chinese) NGP Z370524 WHITESTAR SIGNATURE PRO System Operator’s Manual CD, Multilingual RX Only – NGP Z370500 Rev.
12 INDEX A Accessories 11-1 Add a Surgeon 4-3 Advanced Control Pedal 1-4, 1-18, 2-5, 10-2 Calibration 4-23 Connections 4-17 Default Settings 4-16 Setup 4-16 Switch Assignment 4-17 Testing 4-23 Threshold Configuration 4-22 Advanced Linear Pedal 1-4, 2-5 Battery Replacement 8-4 Connections 4-25 Handle 2-7 Setup 4-25 Switch Assignment 4-26 Test 4-32 Threshold Configuration 4-30 Anterior chamber too deep 8-9 Anterior Segment Surgery 3-2 Aspiration Specifications 10-4 Aspiration Rate Occlusion 5-24 Assign Orde
WHITESTAR SIGNATURE PRO System 12 • Index Connections Advanced Control Pedal 4-17 Front Panel 3-2 Rear Panel 3-3, 4-8 Console Dimensions 10-2 Continuous Burst 5-13 Phaco Power 5-15 Continuous Irrigation 1-5, 3-11 Cup Fill 3-10 Continuous Phaco Power 5-14 Copy a Program 5-29 Create Surgeon 5-2 Cup Fill 3-10 Custom Duty Cycles 5-16 Cut Mode Vitrectomy 5-12 Cut Rate Vitrectomy 5-12 Cutter Vitrectomy 3-7 Diathermy 1-7, 1-10 Default Settings 5-12 Forceps 3-6 Handpiece 3-6, 7-2 Output Power 10-9 Diathermy Hand
WHITESTAR SIGNATURE PRO System 12 • Index Foot Pedal 1-4 Advanced Control Pedal 1-18 Calibrate 4-23 Connections 4-17 Default Settings 4-16 Setup 4-16 Switch Assignment 4-17 Testing 4-23 Threshold Configuration 4-22 Advanced Linear Pedal 2-5 Configuration 4-30 Connections 4-25 Setup 4-25 Test 4-32 Feedback 4-7 Modes Vitrectomy 5-6 Part Number 11-4 Pitch 2-7 Pitch Regions 4-7 Positions 2-7, 5-16 Single Linear 1-4, 2-5 Calibration 4-14 Default Settings 4-7 Setup 4-6 Switch Assignment 4-9 Testing 4-14 Thresho
WHITESTAR SIGNATURE PRO System 12 • Index IV Pole Does not work 8-8 Programmable 1-10, 2-8 Setup 3-3 Testing 6-3 Monitor Drape 2-3 Multiple Burst Phaco Power 5-15 Power 5-13 Multiple Mode Programming 1-6 K N Kick 5-19 No aspiration 8-10 No diathermy or poor diathermy 8-13 No phacoemulsification 8-12 No sound when using diathermy 8-13 No vitrectomy cutting 8-13 L LAMINAR Flow 3-6 Language Selection 6-14 Locked Program 5-28 Long Pulse Phaco Power 5-14 Power 5-13 Low batteries 8-8 Low Power Limit 5-19
WHITESTAR SIGNATURE PRO System 12 • Index Passive Reflux 5-28 Pencil Probe Diathermy 3-6 Percent Kick High End of Range 5-19 Low End of Range 5-19 Peristaltic Pump 10-4 Phaco Default Settings 5-8 Handpiece 3-4 Phaco Handpiece 1-4, 1-10, 3-4, 3-5, 7-2 Part Numbers 11-2 Phaco Power Continuous 5-14 Graph 10-10 Phacoemulsification 1-2, 1-6, 1-9, 3-4, 8-12 Physical Specifications 10-2 Pitch 2-7 Pitch Regions Foot Pedal 4-7 Poor aspiration 8-10 Poor phacoemulsification 8-12 Positions Foot Pedal 2-7 Power Contin
WHITESTAR SIGNATURE PRO System 12 • Index S Screen, Touch Calibrate 6-11 Selecting a Language 6-14 Self Test 3-8 Service Interval 6-18 set vacuum units 1-7 Settings Occlusion Mode 5-21, 5-25 Post Laser Cataract 5-10 Setup Advanced Control Pedal 4-16 Surgical Media Center (SMC) 6-15 System 3-2 Wireless Remote Control 6-5 Short Pulse Phaco Power 5-14 Power 5-13 Shutdown Sequence 3-20 System 3-20 Side vitrectomy 5-6 Single Burst Phaco Power 5-15 Power 5-13 Single Linear Foot Pedal Calibration 4-14 Default Se
WHITESTAR SIGNATURE PRO System 12 • Index Threshold Vacuum 5-24 Threshold Configuration Advanced Control Pedal 4-22 Advanced Linear Pedal 4-30 Single Linear Foot Pedal 4-12 Time Set 6-16 Timers Reset 3-14 Touch Screen Calibrate 6-11 Cleaning 7-2 Does not respond 8-8 Using 3-16 Tubing Cassette 2-3 Install 3-9 Tune 3-11 Begin 3-13 Venturi pump 5-27 View Software Versions 6-12 Vitrectomy 1-7, 1-10, 3-7 Advanced Control Pedal Modes 4-17 Cutter 1-10 Default Settings 5-12 Foot Pedal Modes 4-7 Handpiece 3-7 Pri
WHITESTAR SIGNATURE PRO System 12 • Index RX Only – NGP Z370500 Rev.