COMPACT INTUITION System 1 + Introduction INTRODUCTION About this Manual About Oversimplification COMPACT INTUITIVE System Advanced Fluids System Accessories Safety Precautions Warnings Symbol Definitions SCP Z370415 Rev.
COMPACT INTUITIVE System I + Introduction About this Manual About Oversimplification Indications for Use COMPACT INTUITIVE System This manual includes information about the design of the COMPACT INTUITIVE System for anterior segment (oversimplification) surgical procedures. This manual includes information about optional system enhancements. Y our Abbott Medical Optics Inc. (AMO) representative can confirm the availability of these features for your system configuration in your area.
COMPACT INTUITIVE System 1 = Introduction Advanced Fluids System Accessories Occlusion Mode Poach You can use the Occlusion Mode Poach to regulate the vacuum rise time that follows the occlusion of the poach tip, without limiting the choice of aspiration rate through an occluded needle. In order to independently control the aspiration rate and the vacuum rise time, it is possible io have a different aspiration rate for an occluded needle and an occluded needle.
COMPACT INTUITIVE System 1 « Introduction COMPACT INTUITIVE System Console ELLIPSE FX Hand piece The ELLIPSE FX poach hand piece is available for use with the COMPACT INTUITIVE System. The ELLIPSE FX hand piece provides both longitudinal and transversal movement. You ean use the hand piece with a straight tip or a curved tip. Foot Pedal ‘The operating modes (Diathermy, I/A, Poach, and Hysterectomy) are controlled by the foot pedal.
COMPACT INTUITIVE System 1 = Introduction Each surgical mode and sub mode has its own distinct screen: « Diathermy « Poach « Irrigation/Aspiration + Preciosity Figure 1.1 — Poach Screen Totem wr. 680 a 00% Additionally, there are screens or sub-screens for: + Prime/Tune + End Case = Program Settings + Surgeon Settings + System Setup Prime/Tune Before the start of each surgical case, the system requires that you run prime or prime/tune. The prime mode incorporates the function of clearing the tubing of air.
COMPACT INTUITIVE System 1 + Introduction COMPACT INTUITIVE Operating Modes Continuous Irrigation Continuous Irrigation is immediately available by way of the touch screen. Surgeon control of continuous irrigation with the foot pedal is an available setting for the foot pedal. You can use continuous irrigation to fill cups prior to primness. You can use the cup fill feature in place of continuous irrigation when you fill a cup. The cup fill feature is only available from the prime/tune screen.
COMPACT INTUITIVE System Safety Precautions 1 + Introduction Irrigation/Aspiration (/A) The Irrigation and Aspiration mode allows for controlled aspiration of cortical material from the eye, while maintaining intramuscular stability by replacing the aspirated material with the irrigation solution. A peristaltic pump provides a predictable and stable aspiration rate. You can achieve complete control with “Aspiration” and “Vacuum”. Note: The vacuum units are available in either Memling or kPa.
COMPACT INTUITION System 1 « Introduction 10. Do not operate the system in a condensing environment. Take care to protect the instrument from fluid sprays or fluid buildup. 11. Do not exceed maximum weight of 25 pounds (11.25 Kg) on the Programmable IV Pole bottle holder. 12. Do not use more than one IV pole extender with the TV pole. 13. If there is no IV pole attached io the system, hang the irrigation fluid container at least 77 em from the patient's eye. 14.
COMPACT INTUITIVE System 1 + Introduction Connecting Hand pieces It is very important that the electrical connectors on the hand pieces arc completely dry prior to connecting them to the system receptacles. You may receive a “Poach Hand piece Error” message if the corrector is wet, Handling the Poach Hand piece The poach hand piece is a very delicate instrument and you must handle the hand piece with EXTREME care.
| | | COMPACT INTUITIVE System 1 « Introduction Warnings Foot Pedal Never handle the foot pedal by its power cord. Dao not place the foot pedal on a wet surface. Wireless Remote Control Battery Management System This device complies with Part 15.19 of the Federal Communications Commission {FCC) Rules.
COMPACT INTUITIVE System 1 + Introduction WARNING: DO NOT attempt to use the system if it fails to perform properly as stated in this manual.
COMPACT INTUITIVE System 1 + Introduction WARNING: Place monitoring electrodes or other types of equipment as far from those of the COMPACT INTUITIVE System as possible. AMO recommends high current limiting devices for the protection of such systems. Do not use needle monitoring electrodes. WARNING: Keep the diathermy cord away from the patient and other hand pieces or leads (for example, monitoring electrodes). Keep unused ACTIVE ELECTRODES away from the patient.
COMPACT INTUITIVE System 1 = Introduction WARNING: Disconnect the power before you service the equipment. WARNING: Remove the power cord from the power outlet when the equipment is not in use. WARNING: Do not obstruct the power outlet so you can readily remove the power cord. WARNING: Not recommended for use in condensing environment. If exposed to condensing environment, allow system to equilateral to typical operating room conditions prior to use.
COMPACT INTUITIVE System 1 + Introduction WARNING: Do not use non-AMP approved products with the COMPACT INTUITIVE System, as this can affect overall system performance. AMO cannot be responsible for system surgical performance if you use these products In surgery. Symbol Definitions The following symbols appear on the COMPACT INTUITIVE System: Symbol Definition This symbol on the power switch indicates Power is ON. This symbol on the power switch indicates Power is OFF.
COMPACT INTUITIVE System 1 + Introduction Symbol Definition 1 Patient applied part is isolated from earth ground. 1 Patient applied part is grounded OR no direct electrical energy is involved. Foot Pedal connection.
COMPACT INTUITIVE System 1 « Introduction Symbol Definition IP 1PX4 is the International Protection code that indicates that the device is protected against splashing water sprayed at all angles. IP IP is the International Protection code that indicates that the device is protected against powerful water jets. CE 0413 Indicates compliance with the Medical Device Directive Indicates the authorized European Union representative.
COMPACT INTUITIVE System 1 + Introduction Symbol Definition & Do not re-sterilize. Symbol for sterile medical devices processed using Ethylene [STERILE | [EC] Oxide (EXON. Do not use if package is damaged. Do not use if the product sterilization barrier or its packaging is compromised. Standard Foot Pedal. Shows the current position of the foot pedal as you press the foot pedal and the activated foot switch. The number changes when the position of the foot pedal changes.
COMPACT INTUITIVE System 1 « Introduction Symbol Definition Save As. Creates a new program with the current settings. Restore Program after to any modifications. Must be done before you save any changes. pi Restore Sub mode after to any modifications. Must be done re before you save any changes. Keyboard, when pressed the virtual keyboard shows on the scree. Exit Settings without saving any changes. Indicates the selected surgeon. When selected goes to the Surgeons and Programs screen.