Smart Neck Massager Model: 4098, 4098(RED), 4353 User Manal Document No.
Content 1. PRODUCT INFORMATION.............................................................................................................. 3 2. LIST OF ACCESSORIES AND DETACHABLE PARTS ............................................................ 3 3. PRODUCT DIAGRAM ........................................................................................................................ 4 4. OPERATION METHOD ....................................................................................................
Smart Neck Massager (Model: 4098, 4098 (RED), 4353) 1. PRODUCT INFORMATION Intended use: this Smart Neck Massager has 3 kinds of electric pulse mode, 15 gear electric pulse force, electrical stimulation in the neck to relieve neck pain. It can also provide a constant temperature of 37oC to 43oC to relieve neck discomfort. Frequently used functions: it is suitable for the adjuvant treatment of cervical pain, numbness, headache and dizziness caused by cervical spondylosis.
Note: Model 4098 (RED), Model 4353 and Model 4098 are the identical except for the colour of enclosure. 3.
4.
At least 30 min required for ME equipment to warm from the minimum storage temperature between uses until it is ready for intended use. At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use.
- Output: DC 5 V; 1 A - Rated input voltage shall not exceed 250 V - Classification of protection against electric shock: Class II - The adapter need to comply with IEC 60601-1, and provide 2 MOPP (two means of patient protection) between primary circuit and secondary circuit.
5. SAFETY PRECAUTIONS 5.1 Precautions The warning signs, warnings, and illustrations shown in this manual are intended to enable you to use the product safely and correctly and to prevent injury to you and others. The device can’t be used with other medical electronic instruments, such as cardiac arresters, artificial heart and lungs, medical electronic devices such as medical electronic devices, electrocardiographs, etc., otherwise it will cause danger.
Do not applied the stimulation across or through the head, directly on the eyes, covering the mouth, on the front of the neck, (especially the carotid sinus) or from electrodes placed on the chest and the upper back or crossing over the heart, etc. Be careful to strangulation due to cables and hoses, particularly due to excessive length. Do not posit the adapter to make it difficult to operate the power plug which uses to isolate the equipment circuits electrically form the supply mains.
If cause allergic reactions due to the use of this product, or the performance of ME equipment has changed, stop using it immediately and consult a doctor. The patient is an intended operator. The patient can perform electronic treatment, change battery of remote control and charging the rechargeable battery under normal circumstances. The patient can clean and maintain the device and its accessories according to the user manual when the device not in use.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.
Keep out of reach of children 15
5.
5.
6. FEATURES According to ergonomic design, aesthetic science, in line with neck curve, with hot moxibustion, low frequency electric pulse, electromagnetic effect, circular traction, four-in-one synchronous treatment to form a high using the latest 3D smart fit technology, according to your neck.
8. TECHNICAL PARAMETERS AND SPECIFICATIONS 8.1 Product specifications Model 4098 Product No. 30152 Item Code 1105050001 Rated Voltage 5V ⎓ Rated Power 5W Battery Capacity 1500mAh Executive Standard IEC60601-1:2012 Model 4098(RED) Product No. 34167 Item Code 1105070004 Rated Voltage 5V ⎓ Rated Power 5W Battery Capacity 1500mAh Executive Standard IEC60601-1:2012 Model 4353 Product No.
8.2 Technical parameters Rated load resistance: 500 Ω Mode Rated pulse duration (µs) Rated pulse repetition frequencies (Hz) Maximum amplitude of output voltage (V) 1 40 ‒ 130 12 ‒ 67 62 2 74 ‒ 120 10 62 3 70 ‒ 138 33 62 Output Waveform: see table above DC component: none Single pulses energy: <10 mJ (@ 1000 Ω) Output r.m.
Use faults and solutions 1) The internal rechargeable battery is low, unable to know, and the controller battery is dead. Please confirm if the host product open button has been pressed. 2) Can’t charge Please check if the USB cable and the host and phone power adapter are normal. 3) Is the massage time too long? It is recommended to take 15 minutes at a time, twice a day.
10. SYMBOLS AND SAFETY SIGN Symbol Description Refer to instruction manual/ booklet Type BF applied part Serial number Manufacturer Date of manufacture IP classification: IP22 The first number 2: Protected against solid foreign objects of 12,5mm Ф and greater. The second number 2: Protected against vertically falling water drops when enclosure titled up to 15º. Vertically falling drops shall have no harmful effects when the enclosure is titled at any angle up to 15ºon either side of the vertical.
11. EMC GUIDANCE Guidance and manufacture’s declaration – electromagnetic emission The No touch Infrared Body Thermometer is intended for use in the electromagnetic environment specified below. The customer of the user of the No touch Infrared Body Thermometershould assure that it is used in such an environment. Emission test Electromagnetic environment – guidance Compliance The No touch Infrared Body Thermometeruse RF energy only for its internal function.
Voltage dips, short interruptions and voltage variations on power supply input lines Not applicable Not applicable Mains power quality should be that of a typical commercial or hospital environment. If the user of the No touch Infrared Body Thermometer requires continued operation during power mains interruptions, it is recommended that the No touch Infrared Body Thermometerbe powered from an uninterruptible power supply or a battery.
Radiated RF 10 V/m IEC 61000-4-3 80 MHz to 2.7 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). 10 V/m Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.
Recommended separation distances between portable and mobile RF communications equipment and the No touch Infrared Body Thermometer. TheNo touch Infrared Body Thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
12.
13. CONTACT INFORMATION Guangdong SKG Intelligent Technology Co., Ltd. Address: Building A, 7 Floor, Yingfeng Business Center, No.8 Yixing Road, Junlan Community Residents Committee, Beijiao Town, Shunde, Foshan, Guangdong, China. Customers service hotline: 0757-29811111 For more information, please visit: www.SKG.