Spark User Manual Computed Radiography Reader Doc No. : TM-901-EN Rev 0.1.1 Jun 2013 Part No. : CR-FPM-31-001-EN 3DISC, FireCR, Quantor and the 3D Cube are trademarks of 3D Imaging & Simulations Corp, South Korea, and its affiliates. All other trademarks are held by their respective owners and are used in an editorial fashion with no intention of infringement.
User Manual FireCR Spark TM-701-EN Contact 815, Tamnip-Dong, Yuseong-Gu, Daejeon, Korea Tel : 82-42-931-2100 Fax : 82-42-931-2299 Website : www.3DISCimaging.com E-mail : info@3DISCimaging.com 3DISC Americas 22560 Glenn Dr, Suite 116 Sterling, VA 20164 USA Tel : 1-703-430-6080 E-mail : info@3DISCimaging.com 3DISC Europe Gydevang, 39-41, 3450 Alleroed, Denmark Tel : 45-88-276-650 E-mail : info@3DISCimaging.
User Manual FireCR Spark TM-701-EN The device complies with Part 15 of the FCC Rules. Operation is subject to the condition that this device does not cause harmful interference. NOTE : This equipment has been tested and found to comply with the limits for a Class B Digital Device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
User Manual FireCR Spark TM-701-EN Warnings and Used Symbols To ensure the safety of patients, staff and other persons, any changes to software and hardware delivered by 3D Imaging & Simulations Corp. may only be made with prior written permission from 3D Imaging & Simulations Corp. Please read the respective manuals of the connected software, such as acquisition and diagnostic software, before starting to use the FireCR Spark system.
User Manual FireCR Spark TM-701-EN General Safety Guidelines All the safety and operating instructions should be read carefully before this device is operated. This device has been designed and tested to meet strict safety requirements applicable to medical equipment, and has been supplied in a safe condition.
User Manual FireCR Spark TM-701-EN Do not look inside of the device. If any of the following conditions occur, unplug the device from the electrical outlet and contact authorized service personnel. z The power cord or power adapter is damaged. z An object has fallen into the device. z The device has been exposed to water. z The device has been dropped or damaged. z The device does not operate correctly when the operating instructions are followed.
User Manual FireCR Spark TM-701-EN Index of contents Chapter 1. Introduction ..................................................................... 9 Chapter 2. Unpacking ......................................................................11 2.1. 2.2. Inspection for Damage ....................................................................11 Identify the Components ..................................................................11 Chapter 3. Setting Up ...............................................
User Manual FireCR Spark TM-701-EN Immunity ..................................................................................................... 27 5.4. Laser Safety Statement .................................................................. 29 Chapter 6. Warranty and Repair Service ........................................ 30 6.1. 6.2. 6.3. 6.4. Standard Warranty.......................................................................... 30 Repair Service ................................................
User Manual FireCR Spark TM-701-EN Chapter 1. Introduction Thank you very much for deciding on FireCR Spark – Computed Radiography Reader. The FireCR Spark System is ideal for a wide range of Computed Radiography examinations in specialty practices such as chiropractors, podiatrists and orthopedists.
User Manual FireCR Spark TM-701-EN Main Features State-of-the-Art Scanning Mechanism FireCR Spark adopts state-of-the-art scanning mechanisms using high beam delivery efficiency optics module to construct its compact and rigid structure. High Throughput Its unique and patented dual direction scanning mechanism enables to improve the efficiency and high throughput. Scanning Resolution User selectable resolution of 100μm and 200μm allows user to make diagnosis on variable purposes.
User Manual FireCR Spark TM-701-EN Chapter 2. Unpacking 2.1. Inspection for Damage FireCR Spark is shipped in a custom designed container to protect the reader from external shock. Before unpacking the reader, inspect the shipping container for damage. In case the container is damaged, notify the shipper immediately. 2.2. Identify the Components Identify each of these components. Part No.
User Manual FireCR Spark TM-701-EN WARNING Use of Power Cord; Type SJT or SVT, min. 18AWG, 3-Conductor, VW-1 125V, min 10A (or 250V, 10A). Max 3.0m long; One end with Hospital Grade Type, NEMA 5-15P for 125V or NEMA 6-15P for 250V. Other end with appliance coupler. “CAUTION Grounding reliability can only be achieved when the equipment is connected to an equipment receptacle marked “Hospital Only” or “Hospital Grade”.
User Manual FireCR Spark TM-701-EN Chapter 3. Setting Up WARNING Unsuitable Installation Sites Locations with excessive humidity or dust Locations subject to high temperature Locations subject to shaking or vibration Locations exposed to considerable electrical or magnetic noise, or other forms of electromagnetic energy Locations with poor heat radiation 3.1.
User Manual FireCR Spark TM-701-EN WARNING Do not place anything on top of the scanner. WARNING This equipment may be interfered or interfere by electromagnetic or other interference. Assure a distance of minimum 1.0 m with neighboring equipment. 3.2. Wall Mount (Option) The FireCR Spark can be mounted on the wall to save space occupancy. When purchasing the wall mount kit, a detailed installation guide and all parts necessary for assembly are provided.
User Manual FireCR Spark TM-701-EN WARNING 3D Imaging & Simulations Corp. is not liable for product damage or personal injury when the user fails to follow the product installation instruction. WARNING Pulling, pushing, or climbing on the system may cause the system fall. 3.3. Identify Important Features Look over the reader and features shown in this section. User will need to know where these features are when user operates the reader in later chapters. 3.3.1. Reader Connecting Part 3.3.2.
User Manual FireCR Spark TM-701-EN 3.4. Computer Requirements 3.4.1. Recommended Requirement Operation System CPU Memory Microsoft Windows XP or 7 Core Duo / Core2 Processor RAM 4GB or more Hard Disk 300GB Free Hard Disk Space Network Video Video Resolution 1Gbps Ethernet 32 bit Color Display 1280 x 1024 3.4.2.
User Manual FireCR Spark TM-701-EN reader. 3. Connect the other end of the cable to the USB2.0 / Ethernet port on the computer. DANGER This equipment is indoor use only and all the communication wirings are limited to inside of the building. WARNING Do not pull out the USB / Ethernet cable during scanning. 3.6.2. Connect the Power Cord 1. Connect the power cord to the reader, located on the rear side. 2. Connect the other end of the cord to a grounded power outlet.
User Manual FireCR Spark 3.6.3. TM-701-EN Installation Report After installation of the reader, fill in Installation Report from (Appendix I) and send to 3D Imaging & Simulations Corp. service department by fax or e-mail. Fax : +82-42-931-2299 E-mail : support@3DISCimaging.
User Manual FireCR Spark TM-701-EN Chapter 4. Operating 4.1. System Specifications Sampling Pixel Pitch Pixel Matrix (IP 35cm x 43cm) Pixel Matrix (IP 24cm x 30cm) Pixel Matrix (IP 18cm x 24cm) Standard 200um High 100um Standard 1750 x 2150 High 3500 x 4300 Standard 1200 x 1500 High 2400 x 3000 Standard 900 x 1200 High 1800 x 2400 Accepted Cassette Size 35cm x 43cm, 25cm x 30cm, 24cm x 30cm, 18cm x 24cm Gray Scale Resolution 16 bit Eraser Embedded Computer Interface USB 2.
User Manual FireCR Spark TM-701-EN 4.2. Operation Conditions Indoor use only Operating Temperature 15°C ~ 30°C (59°F ~ 86°F) Temperature Gradient 0.
User Manual FireCR Spark TM-701-EN 4.3. Operating Instructions 4.3.1. Turn on the Reader Turn on the reader. Power switch is on the rear side of the reader. DANGER This device uses laser. Avoid looking inside of the scanner. 4.3.2. Turn on the Computer Turn on the computer. Acquisition and Diagnostic Software must be installed before operate the reader. 4.3.3.
User Manual FireCR Spark 4.3.5. TM-701-EN Circuit Functions Figure 11. Circuit diagram 9 9 9 9 9 9 Power Board : Power board generated DC Power of +15V, -15V, +5V Main Board : This parts controls peripherals for image acquisition. And it delivers amplified digitized signal to PC via USB and Ethernet.
User Manual FireCR Spark TM-701-EN Chapter 5.
User Manual FireCR Spark TM-701-EN Handle with care This side up Non-ionizing electromagnetic radiation FCC ID : X68CRSCANNER3 FCC Mark Medical Equipment WITH RESPECT TO ELECTRIC SHOCK FIRE, AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL60601-1 / CAN / CSA CSS.2 No. 601.1 3SE3 CE Mark 5.1. Manufacturer’s Declaration - Electromagnetic Emission The FireCR Spark system is intended for use in the electromagnetic environment specified below.
User Manual FireCR Spark TM-701-EN 5.2. Manufacturer’s Declaration - Electromagnetic Immunity The FireCR Spark system is intended for use in the electromagnetic environment specified below.
User Manual FireCR Spark Radiated RF IEC 61000-4-3 TM-701-EN 3 V/m 80.0 MHz to 2.5 GHz 3 V/m 80.0 MHz to 2.5 GHz Recommended separation distance Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as deter-mined by an electromagnetic site survey, (a) Should be less than the compliance level in each frequency range (b).
User Manual FireCR Spark TM-701-EN 10 3.70 3.70 7.37 100 11.70 11.70 23.30 For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
User Manual FireCR Spark Radiated RF IEC 61000-4-3 TM-701-EN 3 V/m 80.0 MHz to 2.5GHz 3 V/m 80.0 MHz to 2.5GHz Field strengths outside the shielded location from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than 3V/m.a Interference may occur in the vicinity of equipment marked with the following symbol: Note 1) These guidelines may not apply in all situations.
User Manual FireCR Spark 5.4. TM-701-EN Laser Safety Statement The Computed Radiography Reader is Certified in the U.S. to Conform to the Requirements of DHHS 21 CFR, chapter 1 Subchapter J for Class I(1) Laser Products, and Elsewhere is Certified as a Class I(1) Laser Product Conforming to the Requirements of IEC 60825-1 : 2007. Class I(1) Laser Products are not Considered to be Hazardous.
User Manual FireCR Spark TM-701-EN Chapter 6. Warranty and Repair Service 6.1. Standard Warranty 3D Imaging & Simulations Corp. warrants its non-consumable hardware products to be free from defects in materials and workmanship. The warranty covers the cost of parts and labor to repair the product. Please keep the shipping container for future use. Products returned to the factory for repair should be properly packaged.
User Manual FireCR Spark TM-701-EN 6.4. Shipping The FireCR Spark is a solidly built system designed to survive shipping around the world. However, in order to avoid damage during shipping, the FireCR Spark must be properly packaged. In general, the best way to package the FireCR Spark is in the original factory container. If this is no longer available, we recommend that user carefully wraps the FireCR Spark in at least 75 mm (3 inch) of foam or bubble pack sheeting.
User Manual FireCR Spark TM-701-EN Chapter 7. Technical Assistance If user has any questions about installing or using the device, contact your 3D Imaging & Simulations Corp representative or your local dealer. 3D Imaging & Simulations Corp. 815, Tamnip-Dong, Yuseong-Gu, Daejeon, Korea Tel : 82-42-931-2100 Fax : 82-42-931-2299 www.3DISCimaging.
User Manual FireCR Spark TM-701-EN completed form signed by customer to: Fax : +82-42-931-2299 E-mail : support@3DISCimaging.