User Manual Dental Computed Radiography Reader Doc No. : TM-801-EN Rev 0.1.2 Nov. 2012 Part No. : CR-FPM-11-001-EN 3DISC, FireCR, Quantor and the 3D Cube are trademarks of 3D Imaging & Simulations Corp, South Korea, and its affiliates. All other trademarks are held by their respective owners and are used in an editorial fashion with no intention of infringement.
User Manual FireCR Dental TM-801-EN Contact 815, Tamnip-Dong, Yuseong-Gu, Daejeon, Korea Tel : 82-42-931-2100 Fax : 82-42-931-2299 Website : www.3DISCimaging.com E-mail : info@3DISCimaging.com 3DISC Americas 22560 Glenn Dr, Suite 116 Sterling, VA 20164 USA Tel : 1-703-430-6080 E-mail : info@3DISCimaging.com 3DISC Europe Gydevang, 39-41, 3450 Alleroed, Denmark Tel : 45-88-276-650 E-mail : info@3DISCimaging.
User Manual FireCR Dental TM-801-EN The device complies with Part 15 of the FCC Rules. Operation is subject to the condition that this device does not cause harmful interference. Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
User Manual FireCR Dental TM-801-EN Warnings and Used Symbols To ensure the safety of patients, staff and other persons, any changes to software and hardware delivered by 3D Imaging & Simulations Corp. may only be made with prior written permission from 3D Imaging & Simulations Corp. Please read the respective manuals of the connected software, such as acquisition and diagnostic software, before starting to use the FireCR Dental system.
User Manual FireCR Dental TM-801-EN General Safety Guidelines All the safety and operating instructions should be read carefully before this device is operated. This device has been designed and tested to meet strict safety requirements applicable to medical equipment, and has been supplied in a safe condition.
User Manual FireCR Dental TM-801-EN Do not look inside of the device. If any of the following conditions occur, unplug the device from the electrical outlet and contact authorized service personnel. The power cord or power adapter is damaged. An object has fallen into the device. The device has been exposed to water. The device has been dropped or damaged. The device does not operate correctly when the operating instructions are followed.
User Manual FireCR Dental TM-801-EN Index of contents Chapter 1. Introduction .................................................................... 9 Chapter 2. Unpacking .................................................................... 10 2.1. 2.2. Inspection for Damage ................................................................... 10 Identify the Components................................................................. 10 Chapter 3. Setting Up .................................................
User Manual FireCR Dental TM-801-EN Chapter 5. Symbols ....................................................................... 28 5.1. Manufacturer’s Declaration ............................................................ 29 5.1. - Electromagnetic Emission ........................................................... 29 5.2. Manufacturer’s Declaration - Electromagnetic Immunity ................ 30 5.3. Guidance and Manufacturer ’s Declaration – Electromagnetic Immunity .....................................
User Manual FireCR Dental TM-801-EN Chapter 1. Introduction FireCR Dental Reader • Rapidly view high-quality digital images • Easy-to-use system increases patient throughput • Full-featured, affordable dental solution Experience the Benefits of Digital Imaging The full-featured Fire CR Dental Reader from 3D Imaging & Simulations Corp. rapidly and affordably delivers high-quality digital images for busy dental practices.
User Manual FireCR Dental TM-801-EN Chapter 2. Unpacking 2.1. Inspection for Damage FireCR Dental Reader is shipped in a custom designed container to protect the reader from external shock. Before unpacking the reader, inspect the shipping container for damage. In case the container is damaged, notify the shipper immediately. 2.2. Identify the Components Open the shipping container and identify each of these components. Part No.
User Manual FireCR Dental TM-801-EN WARNING Use of Power Cord; Type SJT or SVT, min. 18AWG, 3-Conductor, VW-1 125V, min 10A (or 250V, 10A). Max 3.0m long; One end with Hospital Grade Type, NEMA 5-15P for 125V or NEMA 6-15P for 250V. Other end with appliance coupler. “CAUTION Grounding reliability can only be achieved when the equipment is connected to an equipment receptacle marked “Hospital Only” or “Hospital Grade”.
User Manual FireCR Dental TM-801-EN Chapter 3. Setting Up WARNING Unsuitable Installation Sites Locations with excessive humidity or dust Locations subject to high temperature Locations subject to shaking or vibration Locations exposed to considerable electrical or magnetic noise, or other forms of electromagnetic energy Locations with poor heat radiation 3.1.
User Manual FireCR Dental TM-801-EN WARNING Sliding the reader may result in internal damage or misalignment of the optics. External vibration or shock during scanning may affect image quality. The reader must be placed on rigid, flat and reinforced desk or tabletop. WARNING Do not place anything on top of the reader. WARNING This equipment may be interfered or interfere by electromagnetic or other interference. Assure a distance of minimum 1.0 m with neighboring equipment. 3.2.
User Manual FireCR Dental 3.2.2. TM-801-EN Touch Display Panel Screen displays reader’s status and control of the reader can be done by touch screen. Touch display panel Figure 3. Touch Display Panel display Status Remark When the reader is turned on, Booting screen booting screen is displayed during system initialization. Power saving mode – IP tray Sleeping moves back into the reader and door will be closed.
User Manual FireCR Dental TM-801-EN Disconnection USB / Ethernet is disconnected. No software Software is not running on PC. Updating Firmware Firmware is being updated. Ready for scan, but no IP Ready for Scan The reader is ready for scan, but IP is not placed on the tray. IP size is recognized and the reader is ready.
User Manual FireCR Dental TM-801-EN Scanning Scanning is in progress. Manual Erasing Erasing image from IP manually. Preview Preview of scanned image. Setting mode Shows information and settings of the reader. Scanning resolution setting 16 Choosing scanning resolution.
User Manual FireCR Dental TM-801-EN Auto sleep setting Choosing time duration until sleep mode. Network setting Manual input for network related settings. DHCP setting Choosing DHCP mode for automatic network settings. IP Address setting Type in IP address manually. Subnet mask setting 17 Type in subnet mask manually.
User Manual FireCR Dental TM-801-EN Default gateway Setting 3.3. Computer Requirements 3.3.1. Recommended Requirement Operation System Microsoft Windows 7 (32 bit or 64 bit) CPU Core Duo / Core2 Processor Memory RAM 4GB or more Hard Disk 300GB Free Hard Disk Space Network 1Gbps Ethernet Video 32 bit Color Display Video Resolution 1280 x 1024 3.3.2.
User Manual FireCR Dental TM-801-EN 3.4. Installation of Acquisition and Diagnostic Software Refer to Acquisition and Diagnostic Software manual. 3.5. Connect the Cable and Power Cord 3.5.1. Connect USB Interface Cable The reader interfaces with computer via USB2.0 cable. 1. Use the USB cable inside the shipping container. 2. Connect the cable to the reader’s USB2.0 port, located on the rear of the reader. 3. Connect the other end of the cable to the USB2.0 port on the computer. Figure 4.
User Manual FireCR Dental 3.5.2. TM-801-EN Connect Ethernet Cable The reader interfaces with computer via Ethernet cable (RJ45 CAT.6 UTP). 1. Connect the cable to the reader’s Ethernet port, located on the rear of the reader. 2. Connect the other end of the cable to the Ethernet port of the Ethernet-hub. 3. If you want to connect to PC directly, use crossed type cable instead of normal Ethernet cable. Figure 5.
User Manual FireCR Dental 3.5.3. TM-801-EN Connect the Power Cord 1. Connect the power cord to the reader, located on the rear side. 2. Connect the other end of the cord to a grounded power outlet. Figure 6. Power Connection DANGER This equipment must only be connected to supply mains with protective earth. Use only a three-wire cord that has grounding. This is a safety feature. If you are unable to insert the plug into the outlet, contact your electrician to replace your obsolete outlet.
User Manual FireCR Dental TM-801-EN WARNING Socket-outlet shall be installed near the device and shall be easily accessible. Do not place the device where difficult to access to appliance inlet. Do not unplug the power cord or turn the power switch off during scanning. 3.5.4. Installation Report After installation of the reader, fill in Installation Report from (Appendix I) and send to 3D Imaging & Simulations Corp. service department by fax or e-mail.
User Manual FireCR Dental TM-801-EN Chapter 4. Operating 4.1.
User Manual FireCR Dental TM-801-EN 4.2. Operation Conditions Indoor use only Operating Temperature 15°C ~ 30°C (59°F ~ 86°F) Temperature Gradient 0.
User Manual FireCR Dental TM-801-EN 4.3. Operating Instructions 4.3.1. Turn on the Reader Turn on the reader. Power switch is on the rear side of the reader. DANGER This device uses laser. 4.3.2. Avoid looking inside of the reader. Turn on the Computer Turn on the computer. Acquisition and Diagnostic Software must be installed before operate the reader. 4.3.3. Imaging Plate Placement and Removal Place the imaging plate on the tray to start scan.
User Manual FireCR Dental TM-801-EN In order to scan or erase the IP, locate the IP on the tray correctly and push the tray into the reader fully until interlock holds the tray. WARNING Locate the IP in correct position. WARNING Do not reuse hygienic bag. 4.3.4. Getting a Scanned Image To getting a scanned image, refer to Acquisition and Diagnostic Software manual. 4.3.5.
User Manual FireCR Dental TM-801-EN Chapter 5. Symbols Symbol Description .
User Manual FireCR Dental TM-801-EN Fragile Handle with care This side up Non-ionizing electromagnetic radiation FCC ID : X68CRSCANNER2 FCC Mark Medical Equipment WITH RESPECT TO ELECTRIC SHOCK FIRE, AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL60601-1 / CAN / CSA CSS.2 No. 601.1 3SE3 CE Mark 5.1. Manufacturer’s Declaration- Electromagnetic Emission The FireCR Dental system is intended for use in the electromagnetic environment specified below.
User Manual FireCR Dental TM-801-EN 5.2. Manufacturer’s Declaration - Electromagnetic Immunity The FireCR Dental system is intended for use in the electromagnetic environment specified below.
User Manual FireCR Dental Conducted RF IEC 61000-4-6 TM-801-EN 3 Vrms 150 kHz to 80 MHz 3 Vrms 150 kHz to 80 MHz Portable and mobile RF communications equipment should be used no closer to any part of the FireCR Dental system, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Radiated RF IEC 61000-4-3 3 V/m 80.0 MHz to 2.5 GHz 3 V/m 80.0 MHz to 2.
User Manual FireCR Dental TM-801-EN a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
User Manual FireCR Dental TM-801-EN 5.3. Guidance and Manufacturer ’s Declaration – Electromagnetic Immunity The FireCR Dental system is intended for use in the electromagnetic environment specified below.
User Manual FireCR Dental 5.4. TM-801-EN Laser Safety Statement The Computed Radiography Reader is Certified in the U.S. to Conform to the Requirements of DHHS 21 CFR, chapter 1 Subchapter J for Class I(1) Laser Products, and Elsewhere is Certified as a Class I(1) Laser Product Conforming to the Requirements of IEC 60825-1 : 2007. Class I(1) Laser Products are not Considered to be Hazardous.
User Manual FireCR Dental TM-801-EN Chapter 6. Warranty and Repair Service 6.1. Standard Warranty 3D Imaging & Simulations Corp. warrants its non-consumable hardware products to be free from defects in materials and workmanship. The warranty covers the cost of parts and labor to repair the product. Please keep the shipping container for future use. Products returned to the factory for repair should be properly packaged.
User Manual FireCR Dental TM-801-EN 6.4. Shipping The FireCR Dental Reader is a solidly built system designed to survive shipping around the world. However, in order to avoid damage during shipping, the FireCR Dental Reader must be properly packaged. In general, the best way to package the FireCR Dental Reader is in the original factory container. If this is no longer available, we recommend that user carefully wraps the FireCR Dental Reader in at least 75 mm (3 inch) of foam or bubble pack sheeting.
User Manual FireCR Dental TM-801-EN Chapter 7. Technical Assistance If user has any questions about installing or using the device, contact your 3D Imaging & Simulations Corp representative or your local dealer. 3D Imaging & Simulations Corp. 815, Tamnip-Dong, Yuseong-Gu, Daejeon, Korea Tel : 82-42-931-2100 Fax : 82-42-931-2299 www.3DISCimaging.
User Manual FireCR Dental TM-801-EN Appendix I Installation Report Please complete this report at the time of installation and submit the completed form signed by customer to: Fax : +82-42-931-2299 E-mail : support@3DISCimaging.