User Manual human Computed Radiography Scanner Rev 0.0.9 February 2010 Part No. : CR-FPM-01-001 ©Copyright 2010 by 3D Imaging & Simulations Corp. All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form or by any means, electronic, mechanical, photocopying, recording or otherwise, without the prior written permission of 3D Imaging & Simulations Corp.
User Manual FireCR Contact 49-3, Moonpyung-Dong, Daedeok-Gu Daejeon, Korea Tel : 82-42-931-2100 Fax : 82-42-931-2299 Website : www.3-disc.com E-mail : sales@3-disc.
User Manual FireCR The device complies with Part 15 of the FCC Rules. Operation is subject to the condition that this device does not cause harmful interference. NOTE : This equipment has been tested and found to comply with the limits for a Class B Digital Device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
User Manual FireCR Warnings and Used Symbols To ensure the safety of patients, staff and other persons, any changes to software and hardware delivered by 3D Imaging & Simulations Corp. may only be made with prior written permission from 3D Imaging & Simulations Corp. Please read the respective manuals of the connected software, such as acquisition and diagnostic software, before starting to use the FireCR system.
User Manual FireCR General Safety Guidelines All the safety and operating instructions should be read carefully before this device is operated. This device has been designed and tested to meet strict safety requirements applicable to medical equipment, and has been supplied in a safe condition. To ensure personnel and patient safety, the device shall be operated and serviced in compliance with all procedures, warnings and precautions during all phases of operation and service of this device.
User Manual FireCR Do not look inside of the device. If any of the following conditions occur, unplug the device from the electrical outlet and contact authorized service personnel. z The power cord or power adapter is damaged. z An object has fallen into the device. z The device has been exposed to water. z The device has been dropped or damaged. z The device does not operate correctly when the operating instructions are followed.
User Manual FireCR Index of contents Chapter 1. Introduction ..................................................................... 9 Chapter 2. Unpacking ......................................................................11 2.1. 2.2. 2.3. Inspection for Damage ....................................................................11 Unpack the Shipping Container .......................................................11 Identify the Components ...............................................................
User Manual FireCR Chapter 5. Symbols ........................................................................ 29 5.1. Manufacturer’s Declaration - Electromagnetic Emission .............. 30 5.2. Manufacturer’s Declaration - Electromagnetic Immunity ................ 30 5.3. Guidance and Manufacturer ’s Declaration – Electromagnetic Immunity ..................................................................................................... 33 Chapter 6. Warranty and Repair Service .......................
User Manual FireCR Chapter 1. Introduction Thank you very much for deciding on FireCR – Computed Radiography Scanner. The FireCR System is ideal for a wide range of Computed Radiography examinations in specialty practices such as chiropractors, podiatrists and orthopedists. Also the FireCR is well suited for small clinics, and mobile units, performing directly at the patient with one compact and affordable package.
User Manual FireCR Main Features State-of-the-Art Scanning Mechanism FireCR adopts state-of-the-art scanning mechanism using swing mirror to construct its compact and rigid structure. High Throughput Its unique and patented dual direction scanning mechanism enables to improve the efficiency and high throughput. Scanning Resolution User selectable resolution of 100μm and 200μm allows user to make diagnosis on variable purposes.
User Manual FireCR Chapter 2. Unpacking 2.1. Inspection for Damage FireCR is shipped in a custom designed container to protect the scanner from external shock. Before unpacking the scanner, inspect the shipping container for damage. In case the container is damaged, notify the shipper immediately. 2.2. Unpack the Shipping Container Open the shipping container by rotating butterfly shape lock and remove items from the container. Butterfly shape lock FIGURE 1.
User Manual FireCR Open the shipping container, and remove the scanner and accessories as described below. Pouch for cassette, manual and software CD Space for power adapter, power cord and USB cable Strap for scanner removal FIGURE 2. Inside of shipping container Pull two straps to remove the scanner. And remove the cassette, manuals and Quantor CD from the pouch attached to the lid.
User Manual FireCR 2.3. Identify the Components Identify each of these components. Item Part No. FireCR Computed Radiography Scanner CR-FP-01-001 Universal Cassette CR-FP-02-001 14” x 17” Imaging Plate (Installed in the cassette) CR-FP-03-001 10” x 12” Imaging Plate (Installed in the cassette) CR-FP-03-002 Quantor Acquisition and Diagnostic Software CR-FP-04-001 USB 2.
User Manual FireCR 14” x 17” Universal cassette with imaging plate Quantor Acquisition and diagnostic software CD USB 2.
User Manual FireCR WARNING Use of Power Cord; Type SJT or SVT, min. 18AWG, 3-Conductor, VW-1 125V, min 10A (or 250V, 10A). Max 3.0m long; One end with Hospital Grade Type, NEMA 5-15P for 125V or NEMA 6-15P for 250V. Other end with appliance coupler. “CAUTION Grounding reliability can only be achieved when the equipment is connected to an equipment receptacle marked “Hospital Only” or “Hospital Grade”.
User Manual FireCR Chapter 3. Setting Up WARNING Unsuitable Installation Sites Locations with excessive humidity or dust Locations subject to high or temperature Locations subject to shaking or vibration Locations exposed to considerable electrical or magnetic noise, or other forms of electromagnetic energy Locations with poor heat radiation 3.1.
User Manual FireCR In case that desk or table is not flat, adjust feet under the scanner for leveling. The scanner weighs approximately 30kg (65 lbs) and requires two persons to lift. Carefully place the scanner into its final position. Adjustable foot FIGURE 2. Scanner leveling DANGER Never place the scanner on the floor. Sliding the scanner may result in internal damage or misalignment of the optics. Install in a location that is level and stable.
User Manual FireCR WARNING Do not place anything on top of the scanner. WARNING This equipment may be interfered or interfere by electromagnetic or other interference. Assure a distance of minimum 1.0m with neighboring equipment. 3.2. Identify Important Features Look over the scanner and features shown in this section. User will need to know where these features are when user operates the scanner in later chapters. 3.2.1. Scanner Connecting Part Power Switch USB 2.0 Port Power Inlet FIGURE 3.
User Manual FireCR 3.2.2. Scanner Status Indication LED Indication LED shows the status of the scanner as described below. LED Green On Remark System On On Blinking On Blinking Status Blue Blinking Cassette is in initial position, and ready to scan. Ready to Scan Scanning & Erasing Cassette is in initial position, and can be ejected. Scan Finished Error Status Indication LED FIGURE 4.
User Manual FireCR 3.3. Computer Requirements 3.3.1. Computer Processor: Intel Core Duo / Core 2 Duo or comparable with AMD Dual Core Processor RAM: minimum 2GByte RAM HD: minimum 80GByte for software and for the archive a hard drive C;\ (for installation) and D:\ (for acquired data, minimum capacity of 25GByte) is preconditioned Network: minimum 100MBit resolution of minimum 1,280 x 1,024 pixel using the true color mode Graphic card / Monitor: for embedded systems a resolution of 1,024 x 768 pi
User Manual FireCR 3.4. Installation of Quantor Please run the included setup “*_setup.exe”. The setup creates a version of the Quantor software on the drive C:\ of your PC. After installation, a Quantor icon will be displayed on the desktop. By double clicking on the icon the Quantor software starts. If the software starts in the demo mode a information displays that the program uses a temporary license. Please confirm this information by clicking on “OK”. The demo license is only available for 20 days.
User Manual FireCR 3.5. Connect the Cable and Power Cord 3.5.1. Connect Interface Cable The scanner interfaces with computer via USB2.0 cable. 1. Use the USB cable in the shipping container. 2. Connect the cable to the scanner’s USB2.0 port, located on the rear of the scanner. 3. Connect the other end of the cable to the USB2.0 port on the computer. FIGURE 5. USB connection DANGER This equipment is indoor use only and all the communication wirings are limited to inside of the building.
User Manual FireCR 3.5.2. Connect the Power Cord 1. Connect the power cord to the scanner, located on the rear side. 2. Connect the other end of the cord to a grounded power outlet. FIGURE 6. Power connection DANGER This equipment must only be connected to supply mains with protective earth. Use only a three-wire cord that has grounding. This is a safety feature. If you are unable to insert the plug into the outlet, contact your electrician to replace your obsolete outlet.
User Manual FireCR WARNING Socket-outlet shall be installed near the device and shall be easily accessible. Do not place the device where difficult to access to appliance inlet. 3.5.3. Installation Report After installation of the scanner, fill in Installation Report from (Appendix I) and send to 3D Imaging & Simulations Corp. service department by fax or e-mail. Fax : +82-42-931-2299 e-mail : service@3-disc.
User Manual FireCR Chapter 4. Operating 4.1. System Specifications Sampling Pixel Pitch Pixel Matrix Scanning Time Standard 200um High 100um Standard 1750 x 2150 High 3500 x 4300 Standard 19 sec High 38 sec Accepted Cassette Size 14 ” x 17” Gray Scale Resolution 16 bit Eraser Embedded Erasing Time 30 sec (User Settable) Scanning & Erasing Cycle Time Standard 49 sec High 68 sec Computer Interface USB 2.0 Dimensions 120 (H) x 460 (W) x 703 (D) mm 4.8 (H) x 18.3 (W) x 27.
User Manual FireCR 4.2. Operation Conditions Indoor use only Operating Temperature 0°C ~ 35°C (32°F ~ 95°F) Temperature Gradient 0.
User Manual FireCR 4.3. Operating Instructions 4.3.1. Turn on the Scanner Turn on the scanner. Power switch is on the rear side of the scanner. DANGER This device uses laser. Avoid looking inside of the scanner. 4.3.2. Turn on the Computer Turn on the computer. Quantor Acquisition and Diagnostic Software must be installed before operate the scanner. 4.3.3.
User Manual FireCR 4.3.5. Circuit Functions FIGURE 7. Circuit diagram 1. Base Board: Base board is a controller of peripherals. It controls peripherals upon command of core board. 2. Peripheral: This is peripherals for image acquisition.
User Manual FireCR Chapter 5. Symbols Symbol Description Warning, Consult Accompany Documents. Manufacturer ON Equipment Power On OFF Equipment Power Off Date of Manufacture FCC ID : X68CRSCANNER FCC Mark Medical Equipment WITH RESPECT TO ELECTRIC SHOCK FIRE, AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL60601-1 / CAN / CSA CSS.2 No. 601.
User Manual FireCR 5.1. Manufacturer’s Declaration - Electromagnetic Emission The FireCR system is intended for use in the electromagnetic environment specified below. The customer or the user of FireCR system should assure that it is used in such an environment Emission test Compliance RF emissions Group 1 CISPR 11 Electromagnetic environment - guidance The FireCR system uses RF energy only for its internal function. Therefore.
User Manual FireCR Power frequency (50/60Hz) Magnetic field IEC 61000-4-8 3.0 A/m 3.0 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. Voltage dips, short <5% Uт (>95% dip in Uт) <5% Uт (>95% dip in Uт) Mains power quality should be Interruptions and for 0.5cycle for 0.5cycle that of a typical commercial or Voltage variations hospital environment.
User Manual FireCR Note 1) Uт is the A.C. mains voltage prior to application of the test level. Note 2) At 80 MHz and 800 MHz, the higher frequency range applies. Note 3) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
User Manual FireCR 5.3. Guidance and Manufacturer ’s Declaration – Electromagnetic Immunity The FireCR system is intended for use in the electromagnetic environment specified below.
User Manual FireCR Chapter 6. Warranty and Repair Service 6.1. Standard Warranty 3D Imaging & Simulations Corp. warrants its non-consumable hardware products to be free from defects in materials and workmanship. The warranty covers the cost of parts and labor to repair the product. Please keep the shipping container for future use. Products returned to the factory for repair should be properly packaged. To obtain warranty service, follow the procedure described in the Repair Service section.
User Manual FireCR 6.4. Shipping The FireCR is a solidly built system designed to survive shipping around the world. However, in order to avoid damage during shipping, the FireCR must be properly packaged. In general, the best way to package the FireCR is in the original factory container. If this is no longer available, we recommend that user carefully wraps the FireCR in at least 75mm (3 inch) of foam or bubble pack sheeting. The wrapped device should then be placed in a sturdy cardboard carton.
User Manual FireCR Chapter 7. Technical Assistance If user has any questions about installing or using the device, contact 3D Imaging & Simulations Corp. Technical Support. 3D Imaging & Simulations Corp. 49-3, Moonpyung-Dong, Daedeok-Gu, Daejeon, Korea Tel : 82-42-931-2100 Fax : 82-42-931-2299 www.3-disc.
User Manual FireCR Appendix I Installation Report Please complete this report at the time of installation and submit the completed form signed by customer to: Fax : +82-42-931-2299 E-mail : service@3-disc.